Approach

SERS-D2DNet transforms spectra from four portable devices toward a Renishaw reference. The study then uses SuperRaman to classify the original and transformed spectra across 20 analyte classes, using 50 spectra per class to train the classifiers.

Reported results

Classification accuracy increased after transformation for all four portable devices in Table 5. The largest gains were for Enwave and Rapid ID.

Enwave accuracy
97.58% → 99.98%
Rapid ID accuracy
96.54% → 98.87%

My contribution

I led the modelling and evaluation of the spectral transformation and recognition workflow, including implementation, generalization analysis, visualization, and writing.

Study scope

These comparisons cover the analytes and acquisition conditions tested in the study. The Zenodo record describes the dataset and how to request access to the restricted files.